17
Specifications (REF: 66801280/66801281)
Vacuum
Continuous Therapy Levels 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg
Intermittent Therapy Levels High: 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg
Low: 0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg
Intermittent Therapy Cycle Times High: 3, 5, 8, 10 minutes
Low: 2, 3, 5, 8, 10 minutes
Alarms
General (all alarms)
Priority Low
Auditory Sound Level Low: 60 dB,
Medium: 68 dB,
High: 74 dB (1 meter from device)
Indicator Color Yellow
Overall alarm delays
Over vacuum less than 5 seconds
High vacuum
Continuous Mode
Intermittent Mode
180 seconds
60 seconds
Low vacuum 60 seconds
Leak 45 seconds
Blockage 120 seconds +/- 5 seconds
Canister full 45 seconds +/- 5 seconds
Low battery immediate
Critical battery immediate
Battery failed 30 seconds
Inactive 15 minutes
Annual maintenance Immediately following startup sequence
Device failed 30 seconds
Power Requirements
Device input voltage 19.5 VDC
Device input power 50 W
Mains Adapter Smith & Nephew REF 66801286
Input: 100-240VAC, 50/60HZ, 10-35VA
Output: 19.5VDC, 2.6A, 50W
Fuses: Internal electronic fuse, not user changeable
Physical
Dimensions 180 x 190 x 76 mm (7" W x 7.5" H x 3" D) with 300ml canister
Weight 1.1 kg (2.4 lbs) with 300ml canister