18
Sound Level
Normal Operations no alarms: <43.7dB
Battery
Operating Time ~ 10-16 hours (therapy) when operating from 25mmHg to 120 mmHg;
~ 8 hours (therapy) at 200mmHg
Type Lithium ion
Compliance UL 2054 / IEC 62133
Safety Protection
Protection Against Electric
Shock
Device internally powered; external power supply. Class II.
Patient Protection Type BF
Ingress Protection IP22 (66801280/66801281)
Environmental
Short term storage and
transport temperature
-25 to 70°C (-13 to 158°F)
Long term storage temperature 5 to 40°C (41 to 104°F)
Operational temperature 5 to 40°C (41 to 104°F)
Relative Humidity 15% to 93% RH
Atmospheric Pressure 700 mbar to 1,060 mbar

Other manuals for smith&nephew RENASYS TOUCH