Ground injection-
The output of the generator is connected to the primary of a Solar Type 6220-2 audio isolation transformer.
The procedure for verifying generator performance is as follows-
Short the secondary of the 6220-2 audio isolation transformer. Using the Solar type 7541-1 spike receptor probe
(20 dB correction factor) and a 9410-1 high voltage attenuator (40 dB correction factor), measure the waveform and
verify it meets the requirement for time to crest and duration.
Calibration loop short