11-7
Caution
Disposable electrodes cannot be reused.
Caution
When installing the thermal recording paper, put the side with grids toward the
thermal print head.
Caution
Use only thermal recording paper supplied by the manufacturer. Using other
paper may shorten the service life of the thermal print head. And the
deteriorated print head may lead to ECG recording of poor quality.
Caution
The device and accessories are to be disposed of according to local regulations
after their service lives.
Caution
When the device is connected to a separate power source other than SUPPLY
MAINS, the separate power source is considered as part of medical equipment.
Caution
When the device is connected to the recorder base, they are considered as a
medical system. While the separate recorder base is non-medical equipment.
Caution
Clinical interpretation of the ECG report may be affected by the filter settings.
Caution
The results given by the device should be examined based on the overall clinical
condition of the patient, and they cannot substitute for regular checking.
12.2.2 Cleaning & Disinfection Cautions
Caution
Turn off the device, disconnect the DC adapter and remove the patient cable
before cleaning and disinfection.
Caution
Prevent the detergent from seeping into the device when cleaning. Do not
immerse the main unit and accessories into liquid under any circumstances.