if the zero offset is ≤±4% then you can use the OFFSET function from the configuration
menu (Keyboard Configuration) to set the zero value. Afterwards, it will be possible to
adjust the translated span, by setting the measurement at URV value using SET S. URV
command:
Initial measurement
The entire span is moved and resized.
With level at minimum
Perform an OFFSET operation. The zero value is adjusted correctly, but the span must be
resized:
With level at maximum
By performing a SET S. URV (set span URV) operation, you can scale the span so that the
current value is resized to be identical to the URV value.