24
Touch to enter the next calibration. Touch can interrupt the calibration.
When the 5 times calibration had been finished or be interrupted, a confirm window will pop up to ask
for save or not save the calibration. (Fig. 5-3)
Touch or to select, then press .
If "YES" is selected, "Calibrate complete!" is displayed.
NOTE:
Set the unit of force to the unit used in calibration previously(as shown in 3.2.1 Unit)
Ensure that the tare weight of attachment has been remove before calibration.
You can do any point or points calibration from 1 to 5 point., we recommend 5 points calibration.