5.2 Calibration
Because of the sensor material performance or the influence of
external factors, there may be errors in a certain range after a
period of time use.
Should send the force gauge to a specialized testing organization
for calibration.
If you have some stansard force weights or the other standard load
and some test stand, you may calibrate it also.
Mount the force gauge.
Remove the tare by use of the key .
Enter Calibration interface, as Fig 5-2.
The calibration interface is shown as Fig 5-3.
Load a standard force.Now the value in
standard input area is just equal to the current measured value.Wait a moment for the force stability.
Touch and to input the stansard force value.
Calibration times
Current measuring value
Standard value input
Fig. 5-3
Fig. 5-4