4 Serial interface
© F. Stephan GmbH SA-112-0106V1.0-WEM-GB 47
cpp5
Start calibration FlowCal=0 PO2Cal=0 PPatCal=35 PRedCal=0
The new values must be saved after calibration!
To this end, adjust YES for the SAVE SETs option in the A
LIA menu.
The new calibration values are only accepted after confirmation by
pressing the IGR!
Saving the values is confirmed on leaving the "Options" menu with the
output:
PCal set
4.2 Setting and querying the date and time
Input of "td" –ENTER starts the output of date and time
td
Date 15:11:22 8.12.05
Input of "td_HH:MM:SS_DD:MM:YY" –ENTER readjusts the time and
date:
ts 15:17:30 10.12.05
The system is set to 10 December 2005, 15.17 hrs and 30 seconds.
The clock in the A
LIA ventilator is not buffered!
It serves merely for time recording when saving interface outputs.