7 Troubleshooting
© F. Stephan GmbH SA-112-0106V1.0-WEM-GB 69
7 Troubleshooting
Alarm text Cause Remedy
O
2
SENSOR FAIL Faulty calibration
Measured sensor voltage outside
the measurable range
Calibrate sensor
Check connection lead
Sensor depleted
CALIBRATION O
2
Faulty calibration
Calibrate sensor
Replace sensor
TEMP.SENSOR Measured temperature close to
patient is outside the measurable
range
Check temperature sensor
Check sensor cable
Replace temperature sensor
Table 3: Troubleshooting table