6 Checks and maintenance
64 SA-112-0106V1.0-WEM-GB © F. Stephan GmbH
Performance and leak test
Siemens lung
PNT D
AIR = 25 l/min
SIMV
Insp = 1 s
Exp = 3 s
Semi-open
PEEP = 3 mbar
PLATEAU = 50 mbar
VT (> 650 ml)
ml
Alarm limits
Insp = 2 s
Exp = 31 s
Alarm
CMV
SIMV A/CV C/BKUP ASB N/CPAP
PEAK PRESS
STENOSIS -
LOW
PRESS/DISCON
-
LOW MIN.
VOLUME
-
High MIN.
VOLUME
-
VT (EXP) LOW -
APNOE -
LED no alarm
HP
Alarm sound HP
MP
Alarm sound MP