6 Checks and maintenance
66 SA-112-0106V1.0-WEM-GB © F. Stephan GmbH
Remove O
2
cell O2 SENSOR FAIL
FiO
2
Criterion
± 3 %
Unit
Results
measuring
unit
Results
Alia
OK
21 20 – 22 %
40 38 – 42 %
60 58 – 62 %
80 77 – 83 %
100 > 97 %
Save settings
Save settings
Pressure measurement
PEEP
PEEP
Criterion
± digit
Unit
Results
measuring
unit
Results
Alia
OK
0 > 1 mbar
2 1 – 3 mbar
5 4 – 6 mbar
10 9 – 11 mbar
Pmax
Pmax
Criterion
± digit
Unit
Results
measuring
unit
Results
Alia
OK
15 14 – 16 mbar
25 24 – 26 mbar
35 34 – 36 mbar