7 Troubleshooting
96
Operator Manual
AMSCO® 2532
10041925_M
General Troubleshooting
Additional Information
Water has not been drained.
1. Press Ack on touch screen to silence buzzer and
acknowledge alarm.
2. Press Resume touch pad to resume cycle.
3. If situation reoccurs, contact STERIS..
7.8 ALARM# 6 SUMP NOT COMPLETELY DRAINED
Condition
Sump filling phase takes less than ten seconds to complete.
General Troubleshooting Additional Information
Water was not completely drained during the previous
drain phase.
1. Press Ack on touch screen to silence buzzer and
acknowledge alarm.
2. Press Abort touch pad to abort cycle.
3. If situation reoccurs, contact STERIS.
7.9 ALARM# 7 SUMP TOO LONG TO HEAT
Condition
Sump water temperature did not reach treatment temperature setpoint within allowed time.
General Troubleshooting Additional Information
Sump water temperature did not reach treatment
temperature setpoint.
1. Press Ack on touch screen to silence buzzer and
acknowledge alarm.
2. Press Resume touch pad to resume cycle.
3. If situation reoccurs, contact STERIS.
7.10 ALARM# 8 SUMP SAFETY WATER LEVEL REACHED
Condition
Defective sump filling valve(s): sump overflow condition detected
NOTE: Sump drain valve opens automatically and closes when overflow condition is no longer detected.