OPERATOR MANUAL
SYSTEM 1
®
endo LIQUID CHEMICAL
STERILANT PROCESSING SYSTEM
CATALOG NO. P6800
S40
®
STERILANT CONCENTRATE
CATALOG NO. S4000
T6800 Rev B

Questions and Answers

  • J
    Jeffrey PenaAug 17, 2025
    What to do if Steris SYSTEM 1 endo shows a power failure warning?
    • W
      walshbrittanyAug 17, 2025
      If your Steris Laboratory Equipment displays a power failure warning, first check the printer by pressing the paper feed button to see if the paper advances. Verify that the Processor was not accidentally unplugged. Inspect all circuit breakers, including the Processor, wall GFCI, and departmental breakers. Then, repeat the cycle. If the issue persists, contact customer service.
  • E
    Edward EdwardsAug 19, 2025
    How to fix water supply problem with Steris SYSTEM 1 endo?
    • R
      Robert ThomasAug 20, 2025
      To address a water supply issue with your Steris Laboratory Equipment, begin by checking the water supply to ensure it is on and sufficiently hot. Inspect the inlet hoses for any kinks and also check the drain hose for kinks. Then, repeat the cycle. If the same message appears, note whether you can open the lid without pressing CANCEL and contact customer service.
  • M
    Matthew JamesAug 22, 2025
    What to do if Steris Laboratory Equipment temperature is over 60 degrees?
    • W
      William EsparzaAug 23, 2025
      If your Steris Laboratory Equipment warns of a temperature exceeding 60 degrees, start by checking the screen in the sterilant compartment for any lint or debris and remove it. Also, check the inlet water temperature; if it's over 60°C, contact your engineering department to have it adjusted. Then, repeat the cycle. If the issue persists, contact customer service.
  • C
    Corey HallAug 26, 2025
    Why does Steris SYSTEM 1 endo take more than 5 minutes to fill?
    • R
      Randall GrahamAug 26, 2025
      If your Steris Laboratory Equipment takes more than 5 minutes to fill, check the water supply to ensure it is on. Inspect the inlet hoses and drain hose for any kinks. Repeat the cycle. If the same message appears, change the pre-filter(s) and run the DIAGNOSTIC cycle. If problems persist, contact customer service.
  • A
    Alvin ObrienAug 28, 2025
    What causes temperature problems in Steris SYSTEM 1 endo Laboratory Equipment?
    • D
      david46Aug 28, 2025
      If your Steris Laboratory Equipment is displaying a temperature problem, check the room temperature to ensure it is not too cold (below 13°C/55°F) or too hot (above 65°C/149°F). Then, repeat the cycle. If the issue continues, contact customer service.
  • C
    Christopher JohnsonAug 30, 2025
    How to resolve a low concentration warning on Steris SYSTEM 1 endo?
    • C
      crystalbenitezAug 30, 2025
      To address a low concentration warning on your Steris Laboratory Equipment, first check the sterilant container to ensure a fresh container is used, that there is powder in the lower section of the cup and that the 'pop-out' plug is still in place. Also, check the tray to see if the screen at the base of the sterilant container is obstructed. Finally, check the Processor to confirm the chamber filled with water and the pumps activated. You can replace the container of sterilant and repeat the cycle. If the issue persists, contact customer service.
  • C
    Cassandra SharpSep 2, 2025
    What to do if Steris SYSTEM 1 endo fill time is less than 0.9 minutes?
    • C
      Carl KerrSep 2, 2025
      If your Steris Laboratory Equipment fill time is less than 0.9 minutes, verify that the MaxLife water filter and pre-filters A and B are installed correctly. Check that the processing tray and/or container is properly in place and that the lid is on. Ensure the device is positioned correctly in the right processing tray or container. Repeat the cycle. If the same message appears, remove the device and any Quick Connect from the processing tray or container and run the DIAGNOSTIC cycle. If the DIAGNOSTIC cycle fails, follow the instructions in Section 8. If the DIAGNOSTIC cycle passes, contact customer service.
  • A
    Amanda HuntSep 5, 2025
    How to fix a chamber open fault on Steris Laboratory Equipment?
    • F
      frank92Sep 5, 2025
      To resolve a chamber open fault on your Steris Laboratory Equipment, check the device and processing tray/container positioning. Look for any material or devices between the top of the processing tray/container and the Processor lid. Then, close the Processor lid and repeat the cycle.
  • B
    bishopduaneSep 8, 2025
    What to do for a heat problem warning on Steris SYSTEM 1 endo?
    • C
      caldwellmichaelSep 8, 2025
      If your Steris Laboratory Equipment shows a heat problem warning, repeat the cycle. If the same message appears, run the DIAGNOSTIC cycle. If the “HEAT RATE LOW” message appears, contact customer service.
  • J
    John HernandezSep 12, 2025
    What to do if Steris Laboratory Equipment chamber temperature is under 45.5 degrees?
    • K
      Kelly PalmerSep 12, 2025
      If your Steris Laboratory Equipment indicates a chamber temperature under 45.5 degrees, run the DIAGNOSTIC cycle. If the “HEAT RATE LOW” message appears, contact customer service.