OPERATOR MANUAL
Reliance
EPS Endoscope Processing System
(2022-03-10) P920011-649

Table of Contents

Questions and Answers

  • N
    Nicholas CarneyAug 11, 2025
    What to do if the Steris Reliance EPS Endoscope Processing System Laboratory Equipment cycle doesn't start?
    • D
      David GreenAug 11, 2025
      If the cycle does not start when you press the START touch pad, make sure the door is completely closed. Push the door firmly to close it, and then press the START touch pad again.
  • G
    Gregory LaneAug 19, 2025
    Why is there little or no water through the Steris Reliance EPS Endoscope Processing System spray arms?
    • P
      perezlisaAug 19, 2025
      There may be little or no water through the rotary spray arms or manifolds because the processor rotary spray arms are clogged or obstructed, the endoscope processing support is not correctly positioned, or the chamber debris screen may be clogged. To resolve this: * Clean the spray arms. * Open the door and firmly push the support into the back of the chamber. * Clean the chamber debris screen if necessary.
  • M
    Miss Nicole DuncanAug 26, 2025
    What to do if the Steris Laboratory Equipment chamber does not drain completely?
    • M
      Michael DavisAug 26, 2025
      If the chamber does not drain completely, the chamber debris screen may be clogged. Clean it if necessary.
  • S
    Shannon CruzSep 3, 2025
    How to turn on the interior light of Steris Reliance EPS Endoscope Processing System Laboratory Equipment?
    • J
      John MeyersSep 3, 2025
      If the interior light does not turn on, it could be because the unit is in the OFF position or the interior light bulb is burned out. Press the top portion of the POWER (ON/OFF) toggle switch to turn on the processor. If that doesn't work, replace the light bulb.

Summary

Overview of Reliance EPS System

Safety Precautions

General Safety Warnings

Covers hazards related to personal injury, equipment damage, electric shock, burn, and health.

Installation Verification

Installation Verification Checklist

Steps to ensure proper installation and operation of the processor.

Component Identification

Processing Instructions

Principles of Operation Overview

Explains the system's high-level disinfection process for endoscopes.

Optional Automated Leak Test

Procedures for setting up and performing automated leak testing on endoscopes.

Processing Cycle Steps

Detailed steps for selecting and running single, double, or decontamination cycles.

Process Verification and Unloading

Procedures for verifying cycle completion and unloading endoscopes.

Faults and Alarms

Guidance on responding to system faults, alarms, and maintenance due messages.

Cycle and Control Value Programming

Supervisor Mode Operations

Configuration settings accessible via Supervisor mode, including cycle values and system options.

Routine Maintenance

General Maintenance Guidelines

Recommendations for frequency and procedures for regular equipment maintenance.

Troubleshooting

Troubleshooting Chart

Common problems and their possible causes and corrections.

Fault Chart

Detailed listing of system faults, their causes, and corrective actions.

Recommended Spare Parts and Consumable Products

Steris Reliance EPS Endoscope Processing System Specifications

General IconGeneral
ModelReliance EPS
TypeEndoscope Processing System
ManufacturerSteris
ApplicationEndoscope reprocessing
Number of endoscopes per cycle1
DisinfectantCompatible with various high-level disinfectants
Processing Capacity1 endoscope per cycle
FeaturesAutomated leak testing
Water ConnectionRequired
Drain ConnectionRequired
Air ConnectionCompressed air supply required
CompatibilityCompatible with a wide range of flexible endoscopes