4-8
920011-649 Operator Manual Processing Instructions
11. If using insertion tube sleeve labeled with a RED dot or a
BLUE dot, proceed as follows (see Figure 4-5):
a. Insert insertion tube sleeve into control handle boot opening,
as shown in Figure 4-5. Verify insertion tube sleeve handle is
pushed down. Insertion tube must be inserted over sleeve
handle to avoid possible damage.
IMPORTANT: Refer to Endoscope Quick Reference Guide
to ensure there are no special placement instructions
required for the endoscope being processed.
IMPORTANT: Verify angulation knobs are unlocked and
distal tip is straight before proceeding to next step.
WAR
NING – INEFFECTIVE
PROCESS HAZARD: When
placing an endoscope in
control handle boot, ensure
endoscope suction barb is
not trapped in the flaps of
universal cord port. Also
ensure light guide cable is
correctly placed between the
flaps of universal cord port.
Figure 4-5. Insertion Tube Sleeves For Small
Diameter Insertion Tubes
Push Down
Sleeve Handle
Flaps
Latch
Insertion Tube
Sleeve With a BLUE
Dot Identification
OR
Insertion Tube
Sleeve With a RED
Dot Identification