4-25
Processing Instructions Operator Manual 920011-649
4.5 Installing Reliance
Dry Chemistry
Container
Disinfection solution is generated within the Reliance EPS.
IMPORTANT: Read Reliance dry chemistry package insert for
detailed instructions for use.
Following directions for use in the Reliance dry chemistry package
insert, prepare and insert container into the processor Reliance CDS.
NOTE: Before inserting Reliance dry chemistry container, verify the
three punching rods inside the Reliance CDS are tight. If not, refer
to S
ECTION 6.7, DRAINING HOLES AND PUNCHING RODS, for procedure.
Once the Reliance dry chemistry container is inserted, close and
latch Reliance CDS.
WARNING – INEFFECTIVE
PROCESS HAZARD:
• Only use Reliance dry
chemistry in Reliance EPS
Processor. Do not substitute
any other product for
Reliance dry chemistry.
• Inspect Reliance dry
chemistry container for
damage prior to use. Do
not use damaged
containers; if damage is
observed, follow directions
for disposal in Reliance dry
chemistry package insert.
• Storage of Reliance dry
chemistry container should
be in a clean dry location.
Reliance dry chemistry
container should not be
exposed to extreme heat or
humidity. Refer to package
insert for instructions.
• Before starting a cycle,
ensure to verify Reliance
dry chemistry container's
expiration date and STERIS
chemical indicator bottle's
expiration date are valid. If
expiration date is not valid,
use a new Reliance dry
chemistry container or a
new chemical indicator
strip. Failure to do so may
result in ineffective
processing.
WARNING – CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
All personnel involved in the
operation of this processor
should be able to demonstrate
the correct procedure for
disposal of Reliance dry
chemistry containers.
Figure 4-22. Reliance Dry Chemistry