4-12
920011-649 Operator Manual Processing Instructions
18. If processing endoscopes without a light guide cable (model
numbers are indicated in Endoscope Quick Reference Guide),
insert a control handle boot plug into universal cord port of
control handle boot (see Figure 4-13).
19. Before closing control handle boot “A”, ensure light guide cable
from control handle boot B will not interfere with proper closure.
20. Close both control handle boots and clasp three latches. Clasp
middle latch first.
Figure 4-12. Plug to Process Only One
Endoscope
CONTROL HANDLE
BOOT “B”
Control Handle
Boot Plug
Fluid
Delivery
Port
Universal
Cord Port
Figure 4-13. Use Control Handle Boot Plug to Process
Endoscope Without a Light Guide Cable
Universal
Cord Port
Control Handle
Boot Plug