7-25
Troubleshooting Operator Manual 920011-649
7.3 Maintenance Due
Alarm Chart
IMPORTANT: When a situation occurs with the Reliance EPS
Processor, an alarm will sound. Refer to S
ECTION 4.13.2,
MAINTENANCE DUE FAULTS, on how to respond to maintenance faults.
MAINTENANCE MESSAGE POSSIBLE CAUSE AND CORRECTION
Alarm sounds, and display shows:
MAINTENANCE DUE
alternating with:
MAINTENANCE 01
SEE CHART
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= MAIN
TENANCE 01 =
DAY COUNTER EXPIRED
1. A general maintenance check is due after a
preprogrammed number of days.**
2. Call STERIS* for a general maintenance check.
** 365 days (default) determined by supervisor and set in
Service mode by service technician (refer to S
ECTION 2,
SERVICE MODE PROCEDURE, of Maintenance Manual
[P764333-736]).
Alarm sounds, and display shows:
MAINTENANCE DUE
alternating
with:
MAINTENANCE 02
SEE CHART
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= MAINTENANCE 02 =
CYCLE COUNTER EXPIRED
1. A general maintenance check is due after a
preprogrammed number of cycles.**
2. Call STERIS* for a general maintenance check.
** 9999 cycles (default) determined by supervisor and set
in Service mode by service technician (refer to S
ECTION 2,
S
ERVICE MODE PROCEDURE, of Maintenance Manual
[P764333-736]).
Alarm sounds, and display shows:
MAINTENANCE DUE
alternating with:
MAINTENANCE 03
SEE CHART
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= MAINTENANCE 03 =
REPLACE AIR PURGE HEPA FILTER
1. Air purge pre-filter and air purge HEPA (High-Efficiency
Particulate Air) filter need to be replaced.**
2. Call STERIS* for a general maintenance check.
** 500 hours (default) determined by supervisor and set in
Service mode by service technician (refer to S
ECTION 2,
SERVICE MODE PROCEDURE, of Maintenance Manual
[P764333-736]).
*Service charges may be incurred. Consult your warranty for details.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris Reliance EPS Endoscope Processing System and is the answer not in the manual?

Steris Reliance EPS Endoscope Processing System Specifications

General IconGeneral
BrandSteris
ModelReliance EPS Endoscope Processing System
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals