4-34
920011-649 Operator Manual Processing Instructions
4.8.4 Low Chemical Warning This option is used to warn operator when level of chemical in Bag-
in-a-Box indicated in display message is getting low. It should
prevent having a chemical empty alarm during a cycle.
Low chemical warning is enabled in Miscellaneous Values menu of
Supervisor mode by setting CHEMICAL WARNING option to YES
(refer to S
ECTION 5.2.3, SETTING MISCELLANEOUS VALUES).
Level of chemical in Klenzyme
®
Enzymatic Presoak and Cleaner
Bag-in-a-Box is automatically verified during start-up of SINGLE and
DOUBLE cycles with WASHING PHASE. Level of chemical in CIP
200
®
Acid-Based Process and Research Cleaner Bag-in-a-Box is
automatically verified during start-up of D-LONG cycle.
1. If chemical level is low when START touch pad is pressed,
display shows:
or:
alternating with:
and:
IMPORTANT: When previous messages are displayed,
approximately five cycles can be processed before indicated
Bag-in-a-Box is completely empty. "FAULT 08 or 09, CYCLE
DELAYED" alarm is triggered, indicating a Bag-in-a-Box is empty.
2. Replace indicated Bag-in-a-Box as explained in S
ECTION 6.8,
R
EPLACING BAG-IN-A-BOX CONTAINER.
or:
Press STOP touch pad to ignore message and process selected
cycle.
KLENZYME
IS LOW
CIP 200
IS LOW
REPLACE AND
PRESS START
PRESS STO
P
TO IGNORE