4-50
920011-649 Operator Manual Processing Instructions
4.14 Shutdown There is no Shutdown mode on the Reliance EPS Processor. The
processor should remain ON at all times except when performing
maintenance or repairs.
If it is necessary to turn OFF the processor, wait until cycle is
completed and in "Ready mode", and display shows:
Press bottom portion of POWER (ON/OFF) toggle switch to turn
processor OFF.
4.15 Power Failure
4.15.1 General 1. If a power failure occurs or if POWER (ON/OFF) toggle switch is
pressed when a cycle is in progress, when power is restored
display shows:
alternating with:
2. Press ALARM REPLY touch pad to acknowledge alarm. Display
shows:
3. Press STOP touch pad to abort. Processor performs Abort
sequence three times: chamber is filled with filtered tap water
and water is recirculated and then drained (approximately
15 minutes total). Display shows:
alternating with:
READY
MM-DD-YY HH:MM XM
WARNING – ELECTRIC SHOCK
AND/OR BURN HAZARD:
Pressing bottom portion of
POWER (ON/OFF) toggle
switch to turn off processor
DOES NOT cut off electrical
power. Lockout/Tagout
building electrical disconnect
switch to turn off power to
processor for maintenance or
repairs.
FAULT 23
CYCLE DELAYED
SEE FAULT CHART
CYCLE DELAYED
STOP TO CANCEL
CANCELLING
CYCLE
PLEASE WAIT
MMM MIN