4-23
Processing Instructions Operator Manual 920011-649
4.4.6 Self-Diagnostic The leak test self-diagnostic is available to the user and is to be
performed without any endoscope to verify the integrity of the
system. The leak test self-diagnostic lasts approximately 90
seconds. Leak test self-diagnostic can be performed with either the
door opened or closed.
Performing such a test without any hose assembly connected in the
processor allows the verification of the system.
To proceed with leak test self-diagnostic without any hose assembly
connected, go directly to Step 5.
R
unning the same test with the hose assembly connected and using
the plug provided on the other extremity allows verification of the
hose assembly. See S
ECTION 4.4.1, PREPARATION, for details on how
to connect hose assembly and see S
ECTION 4.4.3, STARTING CYCLE.
To run leak test self-diagnostic using a hose plug on hose assembly,
proceed as follows:
1. Select appropriate leak test hose (see S
ECTION 4.4.1,
P
REPARATION, and Figure 4-16 for details).
2. Connect quick connect extremity of leak test hose(s) to flexible
connection(s) labeled 1 and/or 2 (see S
ECTION 4.4.1,
P
REPARATION, for details).
3. Select appropriate hose plug.
For example, hose plugs are identified "O" for Olympus
®
endoscopes, "P" for Pentax
®
endoscopes or "F" for Fujifilm
®
endoscopes(see Figure 4-21).
4. Connect hose plug to leak test hose other extremity.
NOTE: Hose plug must remain in place on the extremity of the
leak test hose for the duration of the leak test self-diagnostic.
CAUTION POSSIBLE EQUIPMENT
DAMAGE: Remove any
automated leak test hoses
present in endoscope
processing support before
performing a D-SHORT or a D-
LONG cycle. Performing a
decontamination cycle (D-
SHORT and D-LONG) with
automated leak test hose
assemblies present may
cause water to enter hoses.
Inspect leak test hose
assemblies for presence of
water prior to processing a
cycle. Water in tubing may
cause damage to endoscopes.
REF: 920-502-707
Figure 4-21. Automated Leak Test Hose Assembly
Plug (Olympus Shown)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris Reliance EPS Endoscope Processing System and is the answer not in the manual?

Steris Reliance EPS Endoscope Processing System Specifications

General IconGeneral
BrandSteris
ModelReliance EPS Endoscope Processing System
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals