5-3
Cycle and Control Value Programming Operator Manual 920011-649
5.2 Supervisor Mode Supervisor mode touch pads are hidden behind plastic control cover
to limit access. Supervisor mode includes following menus:
• Change Cycle Values – Includes displays to modify phase
values (see Table 5-1, Cycle Description Chart, for adjustable
phase values).
IMPORTANT: To help ensure safe and effective operation and
process integrity, some preset cycle values cannot be modified.
• Miscellaneous Values – Includes displays to:
- set language selection
- set current time and date
- activate/deactivate bar code reader (accessory), bar code
IDs displays and bar code ID printouts
- set
0.2 μm bacterial-retentive filter countdown
- activate/deactivate WASHING select option
- activate/deactivate repeat WASHING phase
- activate/deactivate water sampling
- activate/deactivate low chemical warning
- activate/deactivate automatic D-SHORT
- select automated leak test feature (if optional Automated
Leak Test Assembly is present)
- set enzyme type
- activate/deactivate printout features (duplicate, multiple and/
or extended printouts)
- print values and cycles
• Prime Pump –
To prime selected cleaner injection pumps
manually.
• Bar Code Test (if accessory is present) – To test proper function
of bar code reader.
• Operator List – If bar code reader accessory is present, to
program bar code Operator IDs in processor control.
• Physician List – If bar code reader accessory is present, to
program bar code Physician IDs in processor control.
5.2.1 Accessing Supervisor
Mode
To access Supervisor mode, proceed as follows (refer to Figure 3-5
for control panel pushbuttons, touch pads and toggle switches):
1. Using a screwdriver, loosen screws and remove control panel cover.
2. Press top portion of POWER (ON/OFF) toggle switch to turn
power ON. Display shows:
TESTING MEMORY
PLEASE WAIT