6-6
920011-649 Operator Manual Routine Maintenance
IMPORTANT: If door is not opened and no touch pad or
pushbutton is pressed once D-SHORT cycle is completed (during
a weekend for example), a second D-SHORT cycle automatically
starts after 36 hours so the processor is ready to use on Monday.
AUTOMATIC D-SHORT feature is set in Miscellaneous menu
(S
ECTION 5.2.3, SETTING MISCELLANEOUS VALUES). AUTOMATIC
D-SHORT screen allows operator to activate or deactivate second
D-SHORT starting after the 36 hours of inactivity.
============================
HH:MM:SS MM-DD-YY
============================
CYCLE - D-SHORT
============================
CYCLE START HH:MM:SS XM
CYCLE DATE MM-DD-YY
CYCLE NUMBER 00000001
UNIT NUMBER 3600000000
============================
PHASE TIME
============================
WASH 1 HH:MM:SS XM
WATER
TYPE HOT
TEMPERATU
RE XX.X
RECIRC. TIME MM:SS
AIR PURGE
TEMPERATURE XXX.X
RECIRC. TIME MM:SS
COMPLETED HH:MM:SS XM
TOTAL CYCLE HH:MM:SS
MAXIMUM
FILLING TIME MM:SS
STOP PRESSED HH:MM:SS XM
SAFTEY DOOR LOCK
DISENGAGED HH:MM:SS XM
============================
= READY TO UNLOAD =
============================
DOOR OPEN HH:MM:SS XM
============================
Figure 6-1. D-SHORT Sample Printout

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris Reliance EPS Endoscope Processing System and is the answer not in the manual?

Steris Reliance EPS Endoscope Processing System Specifications

General IconGeneral
BrandSteris
ModelReliance EPS Endoscope Processing System
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals