6-5
Routine Maintenance Operator Manual 920011-649
7. Close chamber door.
8. Press DECONTAM touch pad once. Display shows:
IMPORTANT:
If a D-SHORT cycle has not been completed within
54 hours, processor control will be automatically locked, i.e. no
SINGLE or DOUBLE endoscope processing cycle can be started
until a D-LONG cycle is performed (see
S
ECTION 6.11, D-LONG
C
YCLE
).
9. Press START touch pad. Display shows:
alternating with:
10. Confirm endoscopes, accessories and Flow Units are not
present within endoscope processing support.
11. Press START touch pad to start D-SHORT cycle. Processor
automatically proceeds through programmed phases of
D-SHORT cycle (see Figure 6-1 for sample printout). After air
purge phase is complete, buzzer sounds and display shows:
then:
Processor cools for 30 seconds. Then display shows:
12. Processor drains for 20 seconds. After draining phase is
completed, buzzer sounds and display shows:
13. Press UNLOCK toggle switch to unlock processor door. Open door.
14. Check debris screen for debris. Rinse screen under running
water if necessary. Rinse screen while it is still wet, before debris
dries. Return screen after cleaning.
D-SHORT
PRESS START
REMO
VE ALL
ENDOSCOPES
PRESS START
WHEN READY
CYCLE COMPLETE
PRESS STOP
COO
LING
PLEASE WAIT
DRAINING
PLEASE WAIT
CYCLE COMPLETE
READY TO UNLOAD