3-3
Component Identification Operator Manual 920011-649
3.1.5 Chamber See Figure 3-3.
The high-quality stainless-steel chamber is insulated to reduce heat
loss and noise level. Four manifold connectors are provided at the
back of the chamber: one is the hot water inlet to the Reliance
™ CDS
Chemical Delivery System, two are connectors for the pressure
sensors and one is for the endoscope processing support. The
endoscope processing support is automatically connected to these
manifold connectors when positioned correctly in the processor
chamber.
A removable debris screen at the bottom of the chamber prevents
large debris from entering the piping system and pump. The screen
is provided with a handle and is easily removed for cleaning.
An interior fluorescent light allows the operator to view inside the
chamber during a Reliance EPS Endoscope Processing Cycle.
3.1.6 0.2 µm Bacterial-
Retentive Filter
See Figure 3-3.
All the water supplied to the processor is filtered through a 0.2 μm
bacterial-retentive filter which is conveniently positioned behind the
front service access panel, below the chamber door, for easy
maintenance.
3.1.7 Cleaner Injection
Pumps
See Figure 3-3.
The peristaltic cleaner injection pumps automatically inject the correct
quantity of cleaner during the appropriate cycle phase. The pumps
are conveniently positioned behind the front service access panel,
for easy maintenance.
3.1.8 Fresh Container
Detection Meter
See Figure 3-3.
The fresh container detection meter is used to detect the presence
of the Reliance dry chemistry container during SINGLE and DOUBLE
cycles.