EasyManua.ls Logo

Steris Reliance EPS Endoscope Processing System - Chemical Additive Descriptions; Washing Detergent; Dry Chemistry

Steris Reliance EPS Endoscope Processing System
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-4
920011-649 Operator Manual Installation Verification
2.3 Chemical Additive
Descriptions
2.3.1 Washing Detergent Klenzyme Enzymatic Presoak and Cleaner Single enzyme-
based detergent for surgical instruments, flexible endoscopes
and other medical apparatus used in healthcare facilities.
This product is provided in a special Bag-in-a-Box container
developed to ease handling and installation.
2.3.2 Dry Chemistry Reliance dry chemistry – When combined with water, this two-
part dry formulation generates a dose of peracetic acid
sufficient to safely and effectively high level disinfect up to two
manually cleaned, immersible reusable, heat-sensitive, semi-
critical devices such as GI flexible endoscopes, bronchoscopes
and their accessories, without damaging the components or
creating hazardous substances. High level disinfection is
ach
ieved within the 50-57°C HLD phase of the endoscope
processing cycle (4-minute generation segment followed by a
6-minute exposure segment).
2.3.3 Decontamination
Chemistry
To help prevent contamination of the processor that may occur during
periods of inactivity, STERIS requires the following product be used
with the D-LONG cycle (refer to S
ECTION 6.11, D-LONG CYCLE):
CIP 200 Acid-Based Process and Research Cleaner A
liquid, acidic detergent system specially formulated to meet
unique cleaning demands. It effectively removes a wide range of
process residues, from inorganic salts to particulate carbon.
This product is provided in a special Bag-in-a-Box container
developed to ease handling and installation.

Table of Contents