4-49
Processing Instructions Operator Manual 920011-649
When Abort sequence is completed, display shows:
and printer records:
==========================
CYCLE CANCEL HH:MM:SS: XX
==========================
IMPORTANT: If a recurring alarm occurs that prevents the Abort
sequence from being completed, call STERIS.
4.13.2 Maintenance Due
Faults
There are four MAINTENANCE DUE alarms indicating when specific
routine maintenance needs to be performed on the processor.
1. When a MAINTENANCE DUE alarm occurs, display shows:
alternating with:
and printer records:
=========================
HH:MM:SS XM MM-DD-YY
=========================
= MAINTENANCE XX =
=========================
2. Press ALARM REPLY touch pad to acknowledge alarm. Display
shows:
NOTE: For MAINTENANCE DUE alarms, refer to S
ECTION 7.3,
M
AINTENANCE DUE ALARM CHART, for a detailed listing of
maintenance alarms. If maintenance is not performed on
processor, MAINTENANCE DUE alarm will be displayed every
24 hours as long as maintenance is not performed and counter is
not reset.
CYCLE CANCELLED
READY TO UNLOAD
MAINTENANCE DUE
MAINTENANCE XX
SEE CHART
READY
MM-DD-YY HH:MM XM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris Reliance EPS Endoscope Processing System and is the answer not in the manual?

Steris Reliance EPS Endoscope Processing System Specifications

General IconGeneral
BrandSteris
ModelReliance EPS Endoscope Processing System
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals