4-48
920011-649 Operator Manual Processing Instructions
Printer records:
=========================
CYCLE RESUME HH:MM:SS XM
=========================
If FAULT alarm occurs during HLD GENERATION or
DISINFECTION phase, the cycle must be aborted (refer to
S
ECTION 7, TROUBLESHOOTING), display shows:
Press STOP touch pad to abort. Processor performs Abort
sequence three times: chamber is filled with filtered tap water
and water is recirculated and then drained (approximately
15 minutes total).
During Abort sequence, display shows:
alternating with:
then:
then:
or:
then:
and printer records:
******************************************
= WARNING =
= LOAD NOT =
= H-L DISINFECTED =
******************************************
STOP TO CANCEL
CANCELLING
CYCLE
PLEASE WAIT
MMM MIN
WARNING
CYCLE CANCELLED
CHEM CONTAINER
MUST BE REPLACED
CHEM CONTAINER
STILL USABLE
LOAD NOT
H-L DISINFECTED
WARNING – INEFFECTIVE
PROCESS HAZARD:
• Correct function of the
system is only ensured after
any diagnosed errors have
been found, corrected and a
successful processing cycle
has been completed.
• Endoscopes and accessories
are not adequately processed,
rinsed and air purged when
processing cycle is aborted.
REPROCESS LOAD.
If Disinfection
Solution Was
Generated Within
Processor
If Disinfection
Solution Was Not
Generated Within
Processor