4-54
920011-649 Operator Manual Processing Instructions
4.17 Disposing of
Expired Dry Chemistry
To dispose of expired Reliance dry chemistry, run a DOUBLE
endoscope processing cycle without washing phase, and without
endoscopes.
See S
ECTION 4.2, GENERAL PROCESSING GUIDELINES, through SECTION
4.7, PRE-PROCESSING CHECKLIST, for additional information.
NOTE: For a 50 Hz processor, refer to the Operator Manual Information
Supplement (P920501-322) for additional information.
IMPORTANT: For this special DOUBLE endoscope processing cycle,
test cords and insertion tube sleeves wi
th a blue dot identification must
be used in both control handle boots "A" and "B". This will maintain
sufficient pressure inside control handle boots in absence of endoscopes.
To run a DOUBLE endoscope processing cycle without washing
phase, and without endoscopes, proceed as follows:
1. Verify display shows:
2. If chamber door is not open, press UNLOCK toggle switch to
unlock processor door and open door. Printer prints:
==========================
DOOR OPEN HH:MM:SS XM
==========================
3. Pull endoscope processing support out of chamber.
4. Open control handle boots "A" and "B". Insert one insertion tube
sleeve with a blue dot identification in insertion tube inlet of
each boot (see Figure 4-27).
READY
MM-DD-YY HH:MM XM
WARNING CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
All personnel involved in the
operation of this processor
should be able to demonstrate
the correct procedure for
disposal of Reliance dry
chemistry containers.
Flaps
Insertion Tube Sleeve
With a Blue Dot
Identification
Figure 4-27. Insertion Tube Sleeve

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris Reliance EPS Endoscope Processing System and is the answer not in the manual?

Steris Reliance EPS Endoscope Processing System Specifications

General IconGeneral
BrandSteris
ModelReliance EPS Endoscope Processing System
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals