8 — Routine Maintenance
19
11042437_A
LED202 In-Ceiling Lighting System Operator Manual
– Quaternary Ammonium Compound (Quats) with IPA + 2-Butoxyethanol solvent
– Quaternary Ammonium Compound (Quats) + Biguanide
– H₂O₂ (Hydrogen Peroxide)
– Neutral Cleaners
– Hypochlorite (3 to 6% solution diluted 1:10 in water)
• Any other protective equipment specified on cleaning and disinfectant product labels.
This procedure should be conducted on a daily basis.
NOTE: Avoid getting water into the internal parts of the wall control. Wring out cloth before wiping any part of the wall
control.
1. Wear rubber gloves (and any other protective equipment specified on cleaning product labels).
2. Rinse all surfaces with a soft cloth wipe and clear water. Wring excess water from cloth before wiping.
3. Prepare cleaning solution according to container directions.
4. Clean surfaces using a soft cloth wipe and cleaning solution. Wring excess solution from cloth before wiping.
5. Rinse all surfaces with a soft cloth wipe and clear water. Wring excess water from cloth before wiping.
6. Wipe all surfaces dry with a fresh cloth.