8 — Routine Maintenance
18
Operator Manual LED202 In-Ceiling Lighting System
11042437_A
8.2 Cleaning the Wall Control
WARNING
PERSONAL INJURY HAZARD.
Customer shall follow the national requirements (standards or directives) for the
hygiene and disinfection for the country in which the equipment is in use.
WARNING
PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD.
Read and follow warnings and instructions found on detergent Material Safety Data
Sheets and labels.
CAUTION
POSSIBLE EQUIPMENT DAMAGE
Always follow manufacturer instructions for concentrations and use of cleaning
products.
Do not scratch optical coating on accessible portions of optical core when cleaning;
always wear rubber gloves and use only a clean, white, lint-free cloth when wiping
internal surfaces.
DO NOT SPRAY any cleaning product directly onto the light engine, wall control or any
system components. Dampen a clean, soft cloth with the cleaning solution and wring
out the excess moisture.
Prevent leakage of fluids into interior of light or wall control. Ensure no excess fluids
remain on light fixture or wall control during and after cleaning.
The following basic supplies are needed to perform routine cleaning of this equipment.
• Pail
• Sponge
• Cloth wipes
• Rubber gloves
• Approved disinfectant cleaner:
– Quaternary Ammonium Compound (Quats) with Ethanol solvent
– Quaternary Ammonium Compound (Quats) with Isopropyl Alcohol (IPA) solvent