ASTOTHERM PLUS Instructions for Use
47
13 Technical data
ASTOTHERM PLUS
REF AP220..
Classification
(IEC/EN 60601-1)
Class I, Type B
applied part is defibrillation proof
Classification
(IEC/EN 60529)
Classification
(MDD 93/42/EEC)
Regulatory class as per FDA
Dimensions
Height
Width
Depth
(incl. attachment device)
max.
145 mm
135 mm
295 mm
Permissible environmental
conditions
10% to 90%
not condensing
10% to 90%
not condensing
Selectable setpoint temperatures
Standard setting by manufacturer
Essential performance acc. to
IEC/EN 60601-1
Regulation of the temperature of the heat
exchanger to a selectable Set-temperature between
37 °C to 41°C +/- 0.5 °C
1.Excessive temperature cut off
2. Excessive temperature cut off
42.5°C (± 0.5°C)
43.5°C (± 0.5°C)
T
Setpoint
– 4°C (± 0.5°C)
Heating up time (22°C to 40°C)
Self-start
after power interruption up to