Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts water which equals 520 ppm available
chlorine (40 mL of a 5.25% bleach solution per 4000 mL water))
70% isopropyl alcohol
1. Make sure that the mattress has been thoroughly cleaned and dried before you apply disinfectants.
2. Wipe down the mattress with a clean, dry cloth to remove any excess liquid or disinfectant.
3. Rinse and dry covers after disinfection.
NNoottee
The mattress cover must be completely dry before you store or add linens. Failure to remove excess disinfectant
could cause degradation of the cover material.
Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If
you do not rinse and dry the product well, a corrosive residue may be left on the surface of the product that could
cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your
warranty.
Frequent or prolonged exposure to higher concentrations of disinfectant solutions may prematurely age the cover
fabric.
The use of accelerated hydrogen peroxides or quaternaries that contain glycol ethers may damage the cover.
0747-109-002 Rev A.3