DDiissiinnffeeccttiinngg
DDiissiinnffeeccttiinngg tthhee pprroodduucctt
WWAARRNNIINNGG
Do not clean, service, or perform maintenance while the product is in use.
Do not steam clean, hose off, or ultrasonically clean the product. Use of these methods of cleaning is not recommended
and may void the product’s warranty.
Recommended disinfectants
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic disinfectant (active ingredient - o-phenylphenol)
Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts water which equals 520 ppm available
chlorine (40 mL of a 5.25% bleach solution per 4000 mL water))
70% isopropyl alcohol
Recommended disinfection method
1. Follow the disinfectant solution manufacturer’s dilution recommendations.
2. Hand wash all surfaces of the product with a disinfectant solution.
3. Avoid over-saturation and make sure that the product does not stay wet longer than the chemical manufacturer’s
guidelines for proper disinfecting.
4. Dry thoroughly. Do not replace the mattress on the product until the product is dry.
5. Disinfect the VVeellccrroo®® after every use. Saturate the VVeellccrroo®® with disinfectant, rinse with water, and allow the
disinfectant to evaporate (appropriate disinfectant is determined by the facility).
6. Check functionality before you return the product to service.
Raise and lower the product
Lock and unlock the brake/steer pedal in both positions
Latch and unlatch the siderails
Raise and lower the Fowler backrest
Raise and lower the gatch
Make sure all components have proper lubrication
Make sure all labels are intact
NNoottee -- Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly.
If you do not rinse and dry the product well, a corrosive residue may be left on the surface of the product that could
cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your warranty.
DDiissiinnffeeccttiinngg tthhee mmaattttrreessss
These instructions provide recommended disinfecting methods for the mattress.
WWAARRNNIINNGG -- Always disinfect the mattress between patients. Failure to do so could result in cross-contamination and
infection.
Recommended disinfectants:
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic disinfectant (active ingredient - o-phenylphenol)
EN 12
0747-109-002 Rev A.3