8001-069-002 Rev B.2 Refer to the Altrix Precision Temperature Management System Operations Manual for additional information.
card 8A
Alarms
Icon Name
Alarm
priority
Message
Therapy
interrupted
Check
Watertemperature
deviation
Medium Watertemperatureis±0.8°C(1.4°F)
outsideoftargettemperature
No Temporaryconditionuponstartup,additionfthermaltransferdevice,or
addition of water
No water Medium No water Yes Check for leaks
Add a minimum of 2 liters of sterile distilled water
Nowaterow Medium Noowdetected Yes Checkforleaksandobstructionsatconnections,hoses,andthermal
transfer devices
Checkwaterowon
anyport
Medium Reducedowdetected No TapConrm,ifthewaterportwasremovedintentionallyCheckforleaks
andobstructionsatconnections,hoses,andthermaltransferdevices
Power loss Medium Notapplicable Yes Checkpowercordconnection
Checkpatientprobe
(AorB)
Medium Abnormalchangeinpatient
temperature
Yes Checkprobecondition,location,andconnections
Probeoradapter
malfunction(AorB)
Medium Notemperaturesignaldetected Yes Checkprobeoradaptercablecondition,location,andconnections
Adaptercable
disconnected(AorB)
Medium Adaptercablenotdetected Yes Reinserttheadaptercable.Ifdamaged,replacetheadaptercable
Patienttemperature
deviation
Medium Patienttemperatureis±0.5°C(0.9°
F)outsideoftargettemperature
No Checkpatientcondition,placementofthermaltransferdevices,andall
connections
Normothermia
deviation
Low
Patienttemperatureisoutsideof
36.0°C(96.8°F)to38.0°C(100.4°F)
No Checkpatientcondition,placementofthermaltransferdevices,andall
connections
Therapypause Medium Therapyiscurrentlypaused Yes Toresume,pressandholdtheplay/pausefor2seconds
Batterylow Low Batteryislow No Maintenanceisrecommended.Ifbatteryisnotreplaced,theproductmay
notfunctiononthenextstartup.
Patienttemperature
output(AorB)
Low Patienttemperatureoutputis
inaccurateontheexternaldevice,
oroutsidesupportedrange
No Checkoutputadaptercableconnection.TapConrmtoreactivatethe
outputport.
Removefromuse
(RFU)
Medium Thesystemhaspoweredoffdueto
a malfunction
Yes Removetheproductfromuseimmediately.Notifytheappropriate
personnel.
Ifanyofthealarmconditionspersist,callmaintenance.
Ifthepageindicatorsappearonthealarmscreen,therearemultipleactivealarms.Onlythehighestalarmisdisplayed.