8001-069-002 Rev B.2 Refer to the Altrix Precision Temperature Management System Operations Manual for additional information.
card 9A
Quick start guide for
Altrix Precision Temperature Management System
Disinfecting
Disinfect the internal water circuit and hoses every 14 days
UsetheBruCleanTbCdisinfectanttabletsbyBrulin(EPAregistration
number71847-2-106)beforerstuse,atleastevery14days,andbefore
storage.BruCleanTbChasbeenvalidatedforinternalwatercircuit
disinfection.Makesurethatyoufollowthedisinfectantmanufacturer’s
guidelinestoavoidtheriskofinjury.Failuretofollowthedisinfectant’s
instructions may void your warranty.
Caution
!
• Alwaysusesteriledistilledwaterorwaterthathasbeenpassed
throughalterlessthanorequalto0.22micronswiththisproduct.
Do not disinfect the internal water system with a thermal transfer
device attached as this may cause a leak.
• Donotusebleachoranyothercleaningordisinfectantagentsfor
internalcircuits.Thiscouldresultindamagetotheproduct.Only
useapproveddisinfectanttablets.
• Alwaysdraintheproductbeforedisinfectingtheinternalwater
circuit.Failuretodraintheproductmayreducetheeffectiveness
ofthedisinfectionprocess.
Tools required:
2gallons(11.4L)ofsteriledistilledwaterorwaterthathasbeenpassed
throughalterlessthanorequalto0.22microns
Personalprotectionequipment(PPE)asrecommendedbythedisinfectant
manufacturer’sinstructions
Soft,lintfreecloth(2ormore)
2BruCleanTbC13.1gtablets(ActiveingredientNaDCCsolutionppm=
1874mg/L)
Note:BruCleanTbCisablendof48%sodiumdichloroisocyanturateand
AdipicAcidwitha5%sodiumdodecylbenzenesulphonatesurfactant.
Servicetooladapterhose(8001-999-017)forColderstyleconnectorhoses
Floor drain
SeeProductillustrationonpage10ofthemanualforclaricationofproduct
componentnamesandlocations.
Draining the internal water circuit and hoses for disinfection
1. Unplugthepowercordfromthewalloutlet.
2. Placethecontrolleroveraoordrain.
Note:Forbestresults,theoordrainshouldbewithinreachofawalloutlettopoweronthecontroller.
3. Todrainthecontroller,pulluponthecontrollerdrainplug(A)toopenthe
drain(Figure24).Leavethedrainopen.
Figure 24
Note: Disinfection
oftheAltrix
internal water
system was
validatedusing
M. mucogenicum.