EENN
www.stryker.com 2971-109-0 01 REV B 1-37
Service
Pod assembly replacement
Tools required:
Zipper pull tool or equivalent
Procedure:
1. Remove the top cover. See Cover replacement, top on page 1-34 for removal.
2. Unsnap the pillow (H) from the left and right side of the pod assembly (Figure 1-12 on page 1-37). Save the pillow.
3. Separate the six Velcr connectors holding the foot section gel crib to the pod assembly.
4. Remove the foot section gel crib (F) by lifting upward (Figure 1-12 on page 1-37). Save the gel crib.
Z
Figure 1-12: Gel crib, pillow, and quick disconnect fitting
5. Disconnect the quick disconnect fitting (Z) (Figure 1-12 on page 1-37).
6. Remove the two retaining pins (A) from the hose quick connection from the foot box assembly (Figure 1-13 on page
1-38). Save the pins.