1-54 2971-109-001 REV B www.stryker.com
Service
Low Air Loss (LAL) fan replacement (Continued)
F
Figure 1-26: Gel crib
12. Remove the two purse lock wire tie (M)(Figure 1-27 on page 1-54). Save the purse lock wire ties.
Figure 1-27: Purse lock wire tie
13. Disconnect the LAL fan from the quick connection.
14. Remove the LAL fan from the LAL fan nest and discard the LAL fan (J) (Figure 1-28 on page 1-55).
Note: During reinstall, route the cable back through the foam and make sure that the new fan is seated into the LAL
fan nest.