16
Safety Shutoff
The L9000 Light Source is equipped with a Safety Shutoff feature which will
temporarily turn off the LEDs in the event of excessive heat in the LED assembly.
W
ARNING ONCE THE LIGHT SOURCE COOLS DOWN (AFTER 7-10 MINUTES),
POWER WILL RESUME TO THE LED MODULE AND THE UNIT WILL
RESTART IN STANDBY MODE. TO PREVENT FIRES AND
ACCIDENTAL BURNS TO THE PATIENT, USER OR INANIMATE
OBJECTS, ALWAYS PLACE THE SCOPES AND/OR FIBEROPTIC CABLES
IN A SAFE PLACE, AND NOT ON THE PATIENT, DRAPES, OR OTHER
FLAMMABLE MATERIAL, TO ENSURE SAFE RESUMPTION OF LIGHT
OUTPUT. IF THE L9000 LIGHT SOURCE EXPERIENCES A
TEMPORARY SHUTDOWN, IT IS RECOMMENDED THAT THE DEVICE
BE RETURNED FOR SERVICE.
Caution Do not abruptly interrupt power to the unit. This will turn off
the fan and may cause severe damage to the internal cooling
system.