Page 4
LIFEPAK 15 Monitor/Debrillator
Noninvasive pacing procedure
Demand pacing
1. Press ON
2. Apply ECG electrodes
3. Apply and connect QUIK-COMBO
®
therapy electrodes;
consider sedation if patient is conscious
4. Press PACER
5. Conrm placement of the sense marker ( )
is near the middle of each QRS
6. Press RATE to select desired pacing rate
7. Press CURRENT until electrical
capture occurs
8. Check blood pressure and pulse to verify
mechanical capture
PAUSE button facilitates temporary viewing
of underlying rhythm
Note: Heart rate alarms are disabled during pacing.