CCoonnttrraaiinnddiiccaattiioonnss
None known.
SSppeecciiffiiccaattiioonnss
WWAARRNNIINNGG -- Always use Stryker approved mattresses that have been tested for compatibility with the product frame to
avoid the risk of patient entrapment.
Safe working load
NNoottee:: Safe working load indicates the sum of the
occupant, accessories, and mattress weight.
1102 lb
500 kg
Maximum patient weight
1027 lb
465.8 kg
Product weight
902 lb
409.1 kg
Scale system capacity maximum 1102 lb 500 kg
Scale system accuracy
± 4.4 lb (2 kg) for patients weighing 225 (102 kg)
to 882 lb (400 kg)
± 6.6 lb (3 kg) for patients weighing 886 (402 kg)
to 1102 lb (500 kg)
Overall length and width
Siderails stowed
92 in. x 38.3 in. 233.7 cm x 97.2 cm
92 in. x 42 in. 233.7 cm x 106.7 cm
92 in. x 48 in. 233.7 cm x 121.9 cm
Siderails at low,
intermediate, and high
position
92 in. x 40.5 in. 233.7 cm x 102.9 cm
92 in. x 46.5 in. 233.7 cm x 118.1 cm
92 in. x 52.5 in. 233.7 cm x 133.4 cm
Patient sleep surface
36 in. (91.4 cm) deck
position
80 in. x 34.5 in.
203.2 cm x 87.6 cm
42 in. (106.7 cm) deck
position
80 in. x 40.5 in.
203.2 cm x 102.9 cm
48 in. (121.9 cm) deck
position
80 in. x 46.5 in.
203.2 cm x 118.1 cm
Bed height to top of seat litter
12 in. to 28 in. 30.5 cm to 71.1 cm
Knee Gatch position 0° to 20° ± 2°
Fowler position
0° to 60° ± 2°
Foot position
0° to 13° ± 2°
Trendelenburg and reverse Trendelenburg
+12° to -12°
Electrical requirements
NNoottee -- Class I Electrical Equipment: Protection against electrical
shock relies on connection to protective earth of an appropriately
rated hospital grade outlet.
120 VAC, 60 Hz, 8A 230 VAC, 50 Hz, 4A
Hospital grade GFCI auxiliary outlet
120 VAC, 60 Hz, 6A
230 VAC, 50 Hz, 3A
EN 8 5900-009-001 Rev A.4