Chaperone
®
Bed Exit System (optional):
Arm/Disarm
The bed exit system is a three-zone center-of-gravity system
that tracks the position of the patient’s body weight and
sounds an alarm when 50 percent of the patient’s body weight
exits a zone.
To Arm
Before arming, make sure the bed has been properly
zeroed without the patient in the bed.
Push and release the Arm/Disarm key. The largest zone
(Zone 1) light will illuminate when armed. The system will
automatically select the largest zone as default.
To change the zone, push the Zone key
until the light indicating the desired zone comes on.
To Disarm
To deactivate Bed Exit, push and hold the Arm/Disarm key.
The selected zone light will turn off.
Note: If the bed was not zeroed at patient admission,
prior to arming, the Chaperone Bed Exit System will not
operate properly.
1
3
3
1
2
2