Secur-Fit Advanced 2 For 1
Tapered Reamer Size 5-6
1601-1556
7 
Reamersareusedtopreparethe
canaldistallywheredensecortical
boneexists,therebyaidingin
broaching.TheTaperedReamer
shouldbeinsertedsuchthatthe
appropriatemarkingcorresponding
tothestemsizecorrelatestothe
depthofthemedialresectionpoint.
See Table 2 for a list of
recommendedreamersizesforeach
stem.
TABLE 2: RecommendedTapered
ReamerSizes
STEM
SIZE
TAPERED
REAMER
4 *Starterreamermaybeused
5 5-6
6 5-6
7 7-8
8 7-8
9 9-10
10 9-10
11 11-12
12 11-12
Size 7
Size 8
TAPERED REAMING
Figure8
If pre-operative templating results
in an in-between size, ream to a
smaller size.
Tip
Secur-Fit Advanced 2 For 1
Tapered Reamer Size 7-8
1601-1578
Secur-Fit Advanced 2 For 1
Tapered Reamer Size 9-10
1601-1590
Secur-Fit Advanced 2 For 1
Tapered Reamer Size 11-12
1601-1512
Due to the aggressive cutting
nature of the reamer, distal
reaming should not exceed the
templated size.
Tip

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker SECUR-FIT ADVANCED and is the answer not in the manual?

Stryker SECUR-FIT ADVANCED Specifications

General IconGeneral
BrandStryker
ModelSECUR-FIT ADVANCED
CategoryMedical Equipment
LanguageEnglish

Related product manuals