STEP 5
TRIAL REDUCTION
10 SECUR-FIT
®
ADVANCED SURGICAL TECHNIQUE
Figure11
Selectanecktrialwhichhasthe
samestemsizeandangleasthe
plannedimplantsize.Thiscanbe
determinedinthreeways.
1.Matchthestemsizeindicatedon
thefaceofthenecktrialtrunnion
withtheinsertedbroach
2.Matchtheneckanglewith
desiredstemanglebasedonthe
127°trialsbeinggold,andthe
132°trialsbeingsilver
3. Matchtheneckangleindicated
onthedistalportionofneckwith
thedesiredstemangle
Thetablebelowindicatesthe
correctnecklengthforeachsize
stem.Thesizeofthebroachdirectly
corresponds to the size of the
implant.
TABLE 3: Secur-FitAdvancedNeck
TrialsandNeckLengths(mm)
Instruments
Secur-Fit Advanced Monolithic
Neck Trial, 127 Degree
(Gold) 1601-XX127
STEM
SIZE
NECK
LENGTH
(mm)
127°
(GOLD)
NECK
LENGTH
(mm)
132°
(SILVER)
4 — 26
5 — 26
6 28 26
7 32 30
8 32 30
9 36 34
10 36 34
11 40 38
12 40 38
Neck
Length
Neck
Angle
Size
30
MM
132
°
Figure12
WARNING
Image depicts
improper use
of the Secur-Fit
Advanced neck trial.
It is important to ensure
the neck trial is placed
properly onto the broach
to avoid improper ROM and
joint stability assessment.
Secur-Fit Advanced Monolithic
Neck Trial, 132 Degree
(Silver) 1601-XX132