C
Courtney StuartJul 25, 2025
What to do if Stryker Water Pump will not turn on?
- KKathryn HernandezJul 26, 2025
If your Stryker Water Pump isn't turning on, there are several potential causes: * Ensure the electrical cord is fully plugged into a properly grounded Hospital Grade receptacle. * Check the power cord for any visible damage and replace it if necessary. * For PC board issues, refer to troubleshooting steps for 'T/PUMP will NOT heat'. * A defective Membrane Panel might be the cause, requiring replacement.