EMC Information
The T/Pump with EN60601-2: 2001 Second Edition
(CISPR Classified as Class B, Group 1 ISM equipment)
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in the following pages.
Portable and mobile RF communications equipment can effect Medical Electrical Equipment.
CAUTION
Do not use the product next to or stacked with other equipment. If next to or stacked is necessary, observe the product
to verify normal operation in the configuration in which it used.
Guidance and Manufacturer’s declaration - Electromagnetic Immunity
The T/Pump system is suitable for use in the electromagnetic environment specified below. The customer or the user
of the T/Pump system should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrostatic fast transient/
burst
IEC 61000-4-4
+2 kV for power supply
lines
+1 kV for input/ output
lines
+2 kV for power supply
lines
+1 kV for input/ output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
+1 kV differential mode
+2 kV common mode
+1 kV differential mode
+2 kV common mode
Mains power quality is that
of typical commercial or
hospital environment.
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
<5%Ut (>95% dip in Ut)
for 0,5 cycle
40%Ut (60% dip in Ut)
for 5 cycles
70%Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
<5%Ut (>95% dip in Ut)
for 0,5 cycle
40%Ut (60% dip in Ut)
for 5 cycles
70%Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the T/Pump system
requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from an
uninterrupted power supply
or a battery.

Table of Contents

Other manuals for Stryker TP700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker TP700 and is the answer not in the manual?

Stryker TP700 Specifications

General IconGeneral
BrandStryker
ModelTP700
CategoryMedical Equipment
LanguageEnglish

Related product manuals