www.stryker.com 101515 REV E 1-19
EMC Information
(Continued)
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note:
U
T
is the a.c. mains voltage prior to applications of the test level.
Recommended separation distances between portable and mobile RF communications equipment and the T/
Pump.
The T/Pump system is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the T/Pump system can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
T/Pump system as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d=1,2√P
80 MHz to 800 MHz
d=1,2√P
800 MHz to 2,5 GHz
d=2,3√P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1
1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
The T/Pump system is suited for use in the electromagnetic environment specified below. The customer or the user
of the T/Pump system should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level Electromagnetic Environment - Guidance