SPECIFICATIONS
FA-20L Compressor
There are two electrical voltage variations:
Compressor P/N Mains Voltage
280504D18N 208-230 VAC +/- 10%, V~ 60 Hz and
280504D27N 220 - 240 VAC +/- 10%, V~ 50 Hz
Electrical Characteristics
Service required: 2 poles, 3 wires (single phase plus protective ground.)
Mains Power Receptacle: Heavy Duty Power Connector, HAN I0E, Harting 09330102602
Power Consumption
Variation Frequency
Voltage Operation
Compressor
Compressor
MCA
MOP
18N 60 220 2.6 12.0
56
15.0 30.0
27N 50 230 2.25 10.0
43-47
13.0 25.0
RLA = rated load current MCA = minimum circuit ampacity
LRA = locked rotor amps MOP = maximum overcurrent protection
Mains Power Supply Cable: 12 AWG, 3 conductor, 300V cordage with Phoenix connector and
NEMA L6-20P (2 pole, 3 wire, 20A, 250 V~) locking plug, P/N 280112C10
Circuit Protection:
Input: 20A, 250V circuit breaker
Control Circuit: (2) FU1 & FU2, 1.0A Slo Blo, 5 x 20 m fuses
Cold Head Valve Motor Circuit: (2) FU3 & FU4, 1.0A Slo Blo, 5 x 20 mm fuses
All fuses are located on front control panel
Class 1: Grounded equipment
Ingress Protection: IP2X
See Compressor wiring diagrams, Figures 20 and 21