EasyManua.ls Logo

Sun Nuclear MapCHECK 3 - Set Up File Manager; Creating a New Patient Plan; Adding Files to Patient Plan; Position MapCHECK 3 for Measurement

Sun Nuclear MapCHECK 3
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Set Up File Manager 9
Deliver Absolute Dose Calibration Measurement
1 Insert MapCHECK 3 into MapPHAN and align to isocenter. See Measurement with MapPHAN
Accessory on page 10.
2 S
et field size to 10 cm x 10 cm (or machine-specific reference field size used for determining
the known dose).
3 In SN
C Patient toolbar, click Load Array Calibration , select the VMAT array calibration
that corresponds to the absolute dose calibration energy, and then click Open. The
default
path to the array calibration files is C:\SNC\SNC Patient\factors\<serial number>.
4 In SN
C Patient menu, select Setup > Calibrate Dose. The Dose Calibration dialog box opens.
See Figure 2-1 on page 8.
5 Follow on-s
creen instructions to collect and save the absolute dose calibration measurement.
If desired, scale the known dose by the linac output at the time of dose calibration.
6 To s
elect this absolute dose calibration as the default, click Set as Default.
7 Click
Close to exit.
Set Up File Manager
The File Manager feature in SNC Patient software is used to store patient data files in a single
location and group them by file type, such as Device Measured, TPS Calculated, DICOM RT Plan,
Images, etc. Each group of patient data files is referred to as a Patient Plan.
When a Patient Plan is selected in the SNC Patient toolbar, commands to load or open a file will
display a list of files in the Patient Plan that match the selected file type.
To access the File Manager feature, select File > File Manager in SNC Patient software or click
FM in the toolbar.
Creating a New Patient Plan
1 Click FM in the toolbar and select New Patient Plan for File Manager. The ‘File Manager:
New Plan’ dialog box is displayed.
2 En
ter Patient ID and Plan ID and then click OK. The ‘File Manager’ dialog box opens.
Adding Files to Patient Plan
1 In the File Manager dialog box, click the heading of the file type to import.
2 In the I
mport dialog box, select the file and then click Open. To select multiple files, press
SHIFT or CTRL while selecting.
The files are copied to the Patient Plan.
3 Click
CLOSE FILE MANAGER.
Position MapCHECK 3 for Measurement
For details about accessories. see See Products, Options and Accessories on page 1.
Measurement Without MapPHAN Accessory
A MapPHAN accessory is not required to collect measurements with the beam normal to the front
surface of the device (e.g., machine QA measurements or collapsed-angle IMRT QA).
As an alternative to the beam straight down setup, the MapCHECK 3 can be attached to the
accessory Gantry Mounting Fixture (GMF) or Isocentric Mounting Fixture (IMF).
1 Pla
ce MapCHECK 3 on treatment couch with +Y toward gantry.
2 A
lign lasers to the MapCHECK 3 X and Y markings. If the area of interest is not centered on
the beam central axis (CAX), such as an asymmetrical field, the MapCHECK 3 can be
positioned off-center. Record the X and Y shift from beam CAX; these values are required to
align to the TPS dose map.

Table of Contents