connected or an applicator is defecve, the display of the iMRS singalizes a failure (Applica-
tor not plugged, Applicator broken). In both cases, you are unable to start the applicaon.
11. Applicator cushion
The applicator cushion can be used for local
applicaons, e.g. knee, foot, hand, shoulder
etc. It is covered with a cover made of medically
approved vinyl (Permaguard®) which can be dis-
infected.
A pair of coils has been integrated into the cus-
hion, therefore it cannot be folded or rolled, but
it can be bent in the middle (do not crease).
Care and Cleaning Guide
The vinyl should be cleaned periodically in order to maintain its appearance and prevent build-
up of dirt and contaminants. Any stain, spills or soiling should be cleaned up promptly to prevent
the possibility of permanent staining.
Please make sure, that the applicator cushion is not covered completely by the body or a blan-
ket or similar when in use. Air circulaon must be guaranteed, otherwise moisture can lead to
corrosion of the coils inside.
The electro-magnec vibraon does not only work directly above the applicator cushion, it vib-
rates in all direcons but prefers to radiate perpendicular to the plane of the applicator cushion.
At the highest intensity seng, the average peak eld strength of the applicator cushion is less
than 40 µT (iMRS wellt) or 70 µT (iMRS complete and iMRS professional) which is approxi-
mately equivalent to the Earth-Magnec-Field. According to the preliminary standard for DIN
0848, the recommended upper limit for 24-hours constant use is 400 µT. Parallel to the plane of
the applicator cushion, the vibraon of the magnec eld has dissipated virtually completely at
a distance of approximately 5/1.5m.
If the applicator cushion is used, the iMRS automacally selects a xed specied program for
local applicaon.
This applicaon can only be used with the original applicator cushion. If no applicator has been
connected or an applicator is defecve, the display of the iMRS singalizes a failure (Applicator
not plugged, Applicator broken). In both cases, you are unable to start the applicaon.